The DiaSpect hemoglobin measuring procedure:
 |
 |
 |
|
1. Make sure the finger is warm and relaxed. Clean and make an insicion in the upper side of the finger tip by use of a suitable minimum 1,5 mm lancette for adults.
|
2. Wipe of the two to tree first drops of blood to get rid of tissue fluid and to get a good blood flow.
|
3. Apply the cuvette in any angle andfill it in one move upwards to avoid possible trapped air bubbles. Make sure to wipe of the remains of blood before possible double sampling.
|
 |
 |
 |
|
4. Make sure there is no blood on the outside measuring surface of the cuvette.
|
5. Place the cuvette in the holder and tap it. The result will be displayed within 1-2 seconds. Make note of the results. Automatic download of results can be done by data management if suitable interface is availible
|
6. In case of a confirmative measurement - within 30-45 seconds - squeeze gently from the finger base to get another drop of blood.
For a possible confirmative measurement; within 30–45 seconds, squeeze gently from the finger base to get another drop of blood.
|
|
|
 |
|
|

|

|
|

|